Volume, Sleep mode, Display brightness – RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 156: Touch panel calibration
![background image](/manuals/466045/156/background.png)
- 156 -
31.4. Volume
The user may adjust power of the loudspeaker which is mounted in a read head:
To adjust the volume:
Press
key, next select „
Other / Volume” option. Using on-screen keyboard
enter percent value of the volume. 100% is the highest volume, 0% the lowest
(loudspeaker off).
31.5. Sleep mode
The user has the sleep mode at his disposal, thus he can turn off the display.
In order to enter the sleep mode:
•
press
key, next press: „
Others / Sleep mode”.
On entering the option one of the following settings must be selected: [None;
1; 2; 3; 5; 10; 15]. Digit values stand for a number of minutes. Upon
selection of a desired value the program returns to the previous menu level.
Caution:
The screen turns to sleep mode only when the balance is not operated (the indication
on the screen remains the same). Upon turning to the sleep mode the balance
returns to weighing mode automatically if only any weight modification is detected or
screen touch or key on an overlay pressed.
31.6. Display brightness
Display brightness influences balance operation time, when the balance is
powered by means of a battery. If the user wants to lengthen the operation
cycles as much as possible he should reduce the brightness of the display.
In order to enter the display brightness:
•
press
key, next press: „
Others / Display brightness”.
On entering the option enter a desired value [0% - 100%]. Upon entering a
particular value the brightness changes respectively and the program returns
to the previous menu level.
31.7. Touch panel calibration
Display calibration is required if during operation a user spots incorrect
reaction of the touch panel.
Procedure:
•
Enter submenu<
Others>
•
Select parameter <
Touch panel calibration> which opens an editing
window,