Working modes, Readout – RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 43

- 43 -
14.2.2. Working modes
On entering this option, the display opens a window containing all accessible
working modes. The user can introduce their settings to each of the working
modes which are activated on selecting a specific profile.
In each of the working modes, the user can change the following parameters:
Settings
contain specific parameters relating to a working mode and universal
settings, such as: result control, tare mode, automatic footer printout,
printout mode, printout.
Functions of quick access keys
declaring quick access keys, which are visible in the bottom display bar
Information
declaring information which is visible in the grey coloured workspace
Printout
declaring type of printout of defining a non-standard printout
14.2.3. Readout
The function enables the user to adjust balance operation to ambient
conditions (filter settings) or individual user needs. Menu
the following elements:
FILTER
Each measurement signal before being displayed is electronically
processed for the purpose of obtaining corrected parameters
specific for stable measurement result, i.e. ready to read.
Balance user can influence to some extend the range of signal processing
by selecting an appropriate FILTER value. Available options:
o
very fast
o
fast
o
average / normal
o
slow
o
very slow
While setting the filtering level the user should consider the actual
operating conditions of a balance. In case of very good operating
conditions set the filter to average/normal or fast, in case of rough
conditions set the filter to slow or very slow.
Caution:
in case of precision balances 3Y series the recommended range of filter
is between very fast
÷
average/normal;
in case of analytical balances and microbalances 3Y series the
recommended range of filter is between average/normal
÷
very slow.