14.2.2. Working modes

On entering this option, the display opens a window containing all accessible

working modes. The user can introduce their settings to each of the working

modes which are activated on selecting a specific profile.

In each of the working modes, the user can change the following parameters:

Settings

contain specific parameters relating to a working mode and universal

settings, such as: result control, tare mode, automatic footer printout,

printout mode, printout.

Functions of quick access keys

declaring quick access keys, which are visible in the bottom display bar

Information

declaring information which is visible in the grey coloured workspace

Printout

declaring type of printout of defining a non-standard printout

14.2.3. Readout

The function enables the user to adjust balance operation to ambient

conditions (filter settings) or individual user needs. Menu contains

the following elements:

FILTER

Each measurement signal before being displayed is electronically

processed for the purpose of obtaining corrected parameters

specific for stable measurement result, i.e. ready to read.

Balance user can influence to some extend the range of signal processing

by selecting an appropriate FILTER value. Available options:

o

very fast

o

fast

o

average / normal

o

slow

o

very slow

While setting the filtering level the user should consider the actual

operating conditions of a balance. In case of very good operating

conditions set the filter to average/normal or fast, in case of rough

conditions set the filter to slow or very slow.

Caution:

in case of precision balances 3Y series the recommended range of filter

is between very fast

÷

average/normal;

in case of analytical balances and microbalances 3Y series the

recommended range of filter is between average/normal

÷

very slow.