Acquiring part mass from database – RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 56
- 56 -
Caution:
Remember:
•
Total mass of all parts loaded on the weighing pan must not exceed the
maximal capacity (weighing range) of the balance;
•
Total mass of all parts loaded on the weighing pan must not be lower
than value declared in parameter “Minimal reference mass”. Unless
this condition is met, the balance displays a message:
mass>;
•
Mass of a single part must not be lower than 0,1 of balance’s reading
unit. unless this condition is met, the balance displays a message:
low part mass>.
16.5. Acquiring part mass from database
A product record in the database of products features set of information which
identifies it. One of them is mass, which is used during part counting process.
Procedure:
When in working mode < Parts counting> press <
Products database>
key, and then select desired product from displayed list.
Insert standard (reference) mass to balance’s memory
Procedure for adding standard (reference) mass of a single part to the
database of products:
a) Press < Setup > key followed by pressing < Database > key
b) In the Database of Products press < Products >
c) Press name of a product and edit data in field no. [5] Mass
d) Return to
If there are no data in the database:
a) Press < Setup > key followed by pressing < Database > key
b) In the Database of Products press <
Add > key
c) Accept the process of adding a new record in the database
d) Fill in field referring to a product, including field no. [5] Mass
e) Return to