RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 87

- 87 -
PROCEDURE 3 – Preparing a formulation that is saved in the Database of
Formulas, with determining target mass of complete formulation.
This procedure is dedicated for users who need to prepare a mixture with
pre-determined mass, that is different from the mass resulting from
summing of used ingredients.
In such case, there is no need to carry out complex calculations of
ingredients’ mass. The software adjusts mass of each ingredient in relation
to pre-defined target mass of the complete formulation.
Follow hints and commands displayed by the balance’s software
•
Press <
Formulation> key,
•
Using list of available formulas, select one for making process,
•
Press <
Formulation with target mass> key,
•
The software displays a window with keyboard for inserting the target
mass of a mixture,
•
On accepting the inserted value, the software proceeds to formula making
process. The grey coloured workspace on the display indicates the
following parameters: name of selected formulation, name of the first
ingredient, ingredient number, mass to be weighed and target mass,
mass of each ingredient is automatically recalculated (proportionally) to
the value of defined target mass of the complete formulation,
•
After accepting mass of an ingredient, the software automatically adds the
mass to the sum of weighed ingredients of the formulation, and moves to
weighing the following ingredient,
•
On weighing all the ingredients the software automatically prints a report
on completed process and saves it in the database of formulation reports
(report template is editable in option
> key to
confirm completing and proceed.
Each report on completed formula making processes can be printed from
Database level – reports from formulas.