Sensor sensitivity, Proximity sensors delay, Autotest – RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 158

- 158 -
31.9. Sensor sensitivity
The parameter is adjustable between 0 – 9, and it determines the
distance from which the IR proximity sensors should react.
As standard, the sensitivity is set within the range of 5 to 7.
Procedure:
•
Enter submenu<
Others>
•
Select parameter
Select one of available options. The selection causes immediate returning to
the menu window.
31.10. Proximity sensors delay
Parameter for setting delay of proximity sensors, it is expressed in [ms].
By default the value is set to <0>.
When using option of motion recognition the user must remember about setting sensors delay> parameter to value <500ms> Procedure: a) Go to < Other> submenu. b) Select window with an on-screen keyboard to open. c) Enter value <500>. d) Confirm. The program returns to menu window. 31.11. Autotest AUTOTEST function is designed to aid a user in both assessing balance’s operation and diagnosing the reasons for occurrence of errors in weighing which exceed the maximum permissible values for a given balance model. By means of a simple, repeatable and fully documented way the function enables optimizing balance’s settings to maintain the best possible repeatability and weighing time at a workstation. The main purpose of function operation is the possibility of monitoring the above mentioned parameters at optional moment and saving records from carried out tests. The functions is divided into two modules: AUTOTEST FILTER AUTOTEST GLP Before carrying out any of the tests , the balance controls its level status, temperature and humidity.