DOZOWANIE
Uses the Target Value expressed in [%], which has to be reached during

weighing (pouring) a sample. The target value also features a [+/-] tolerance,

determining the area within which the measurement is recognized as correct.

PROCEDURE:

1. Press <

Target Value > key,

2. Insert the target value expressed in [%],

3. If tolerance applies, set its value,

4. Accept by pressing <

> key,

5. Press High Threshold key and insert its value in [%],

6. Accept by pressing <

> key

19.3. Interpreting the function by use of a bargraph

The dosing and checkweighing functions are aided by graphic indication, in a

form of a bargraph. Below see an example of simultaneous operation of the

two functions.

a) checkweighing thresholds<

> are set as Low Threshold = 90%,

High Threshold = 110%

b) Target value = 105% ; tolerance = 5% <

>

c) Reference mass = 19.986 g <

>

19.4. Additional settings of Percent Setup mode

The additional settings enable adjusting the working mode to user’s needs and

requirements. Functions related to checkweighing: result control, tare mode,

automatic footer printout, printout mode / value release, printouts.

Means of using the above functions are given in point 10.8. “Additional

parameters on weighing process”.