Messages displayed during a control process:
Batch quantity 1
/ 10
- A command on process course and the quantity
of all measurements in a tested batch
Product
- Name of the controlled product
Average
- Average mass of the controlled product
Standard
deviation
- Standard deviation from a controlled batch
After completing the control process, the software generates a report, a the
completed control is automatically saved in the corresponding database.
Caution:
A template and an example of a report on a control process is provided in point
26.7 of this user manual.
26.4. Report on product control
An example of report on product control:
----------------- SQC ------------------
User
Product 000
Start date 2012.10.19 12:27:09
End date 2012.10.19 12:37:20
Batch number ppp
Batch quantity 10
Nominal mass 50 g
Limit T2- 5 g 10 %
Limit T1- 2.5 g 5 %
Limit T1+ 1 g 2 %
Limit T2+ 2 g 4 %
--------- Measurement 1 --------------
Net 49.998 g
--------- Measurement 2 --------------
Net 49.998 g
--------- Measurement 3 --------------
Net 50.021 g