Minimum sample – RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 29

- 29 -
−
Printout
Contains type of printout that is related to a working mode. Printout
takes place on pressing PRINT key on balance’s overlay.
Accessible options:
•
STANDARD PRINTOUT
Possible declaration of printout content: HEADER, WEIGHING DATA
and FOOTER.
printout components marked as
pressing printout activating function key.
•
NON-STANDARD PRINTOUT
Database of printouts enables selecting one of available non-
standard printouts visible in menu
unique printout which is automatically added to the databases.
Caution: Means of designing printouts is described in point 15.5 of this
user manual.
−
Air buoyancy compensation
Contains parameters allowing the user to switch on the correction and
enter data relating to density of the sample and to density of the air.
Caution: function operates only for weighing mode.
For instruction and information on air buoyancy compensation operation
see the previous section of this manual.
10.8. Minimum sample
The weighing mode settings comprise a function of
function is enabled on specifying in the database sample> parameters on methods for determining the minimum sample and the values of minimal sample weight specific for a method. In standard version, this database item is not filled in with data. Activities related to determining minimum sample and adding data to the database: an authorized “RADWAG” personnel. Should balance operator use the minimum sample mode, and data on minimal sample weight are not entered to balance menu, then please contact the local representative of “RADWAG” company. An authorized RADWAG representative determines the minimum sample dependent on packaging mass. The process is carried out with mass standards, in balance place of use and in accordance with the requirements of the accepted quality system. The obtained values are entered to balance software: A minimum sample determining method may have a few tare values defined, along with assigned values of minimum samples, expiry dates of the carried out measurements and entered data. The settings are not to be edited by balance user.