If value greater than the valid one is given, the BEEP sound is activated for
the maximal available time limit.
Example
Command:
BP_350
Response:
BP_OK
Caution! – BEEP signal activated by BP command gets aborted if it is activated
by means of other devices: keyboard, touch panel, proximity sensors.
OMI - list available working modes
The command lists available working modes for a given device
Format: OMI
Accessible responses:
OMI
n_”Mode name”
:
n_”Mode name”
OK
– command completed, available working modes are
given
OMI_I
– command understood but not accessible at this
moment
Mode name – parameter, working mode name, given inbetween inverted
commas. Form of the name is the same like on the balance display, it is given
in a currently operating language.
n – parameter, decimal value specifying working mode number.
n → 1 – Weighing
2 – Parts counting
3 – Percent setup
4 – Dosing
5 – Formulation
6 – Animal weighing
7 – Density
8 – Density of solids
9 – Density of liquids
10 – Peak hold
11 – Totalizing
12 – Checkweighing
13 – Statistics
14 – Pippette calibration
Caution: working modes numbering is assigned to working modes names and
it is the same for each balance types.
Some balances in response to OMI may give working modes numbers without
names.