RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 44

- 44 -
Value release
This parameter determines displaying the stability marker for a
measurement result.
There are 3 available settings of value release parameter
o
fast
o
fast + reliable /recommended/
o
reliable
Caution:
The speed of stability marker occurrence depends on applied filter value and
applied value release.
Autozero function
The function is to automatically monitor and correct zero
indication of a balance.
If the function is enabled, the following measurement results are
compared to each other in constant time intervals. If the results differ
less than declared AUTOZERO range, e.g. 1 division, the balance is
automatically zeroed, and the markers of stable indication
and
precise zero
are displayed.
Active AUTOZERO means, that each measurement starts from the precise
zero point. There are, however, cases where the function may disturb the
weighing process. For instance during very slow load placing or pouring
onto the balance’s weighing pan. In such case, the correcting system of
zero indication may also correct the actual indication of a load placed on
the weighing pan.
Accessible settings: NO - autozero function disabled
YES - autozero function enabled
Autozero function: Dosing
The function sets autozero operation mode to default one set for
dosing mode.
Accessible values:
NO - autozero operation is automatically disabled on entering
dosing mode
YES - autozero operation is automatically enabled on entering
dosing mode
Last digit
The function determines visibility of the last decimal place
indicated on measurement result. The functions provides three
settings:
- ALWAYS: al the digits visible
- NEVER: last digit invisible
- WHEN STABLE: last digit visible only for a stable
measurement