Procedure:

Connect a transponder card scanner to a balance’s port (RS 232 COM1 or

RS 232 COM2), select communication port in balance settings for the

transponder card scanner

In submenu <

Communication> set baud rate for compatible with the

one of the transponder card scanner (default 9600b/s).

Enter the database of operators and edit desired operator’s record, go to <

Card number> option,

On entering editing mode in <

Card number> option, an editing window

is opened with an on-screen keyboard,

Approximate the card to the transponder card scanner. The balance

automatically displays number of detected card in the

field,

Accept recognized number by pressing

key and returning to the main

window of weighing mode.

29.5. Additional display

The balance enables communicating with an additional display using the

following ports:

RS 232 (COM1),

RS 232 (COM2),

Tcp.

Type of data displayed on an additional display is declared in variable

(template)>.

Procedure – designing a variable

Enter group of parameters <

Peripheral devices>

Press “

Additional display /

Project” field, which opens an editing

window with an on-screen keyboard,

Use the on-screen keyboard to insert desired value of a standard, or

Select desired value of a standard using a list displayed on pressing <

> key

Accept by pressing

key.

Caution:

Balance 3Y series cooperates with an additional display WD5/3Y manufactured by

RADWAG.

Ensuring correct cooperation between the balance and the additional display, requires
that the parameter <

Standard> has a variable number set to value {140}, and

the baud rate value set to 38400 bits/s for a port having the additional display

plugged.

29.6. Ambient conditions module

A balance 3Y series enables connecting an ambient conditions module THB

series using ports COM 1 or COM 2.