Pipettes calibration – quick access keys, Adding a pipette to the database – RADWAG MYA 5.3Y.F1 Microbalance User Manual

Page 97

background image

- 97 -

3. Press key,

4. The display indicates functions related to calibration of pipettes.

Pipettes calibration is related to the following functions:

Number of measurements; this option enables determining number of

measurements for tested volume of a pipette (valid for each tested

volume in case of a pipette with adjustable volume),

Request serial number: YES/NO. For option before starting

calibration procedure, the software displays a window and on-screen

keyboard for inserting serial number of a pipette,

Comply with ISO 8655: YES/NO. For option , for tested

volumes, during pipettes calibration process the software automatically

accepts error values as specified in standard ISO 8655 (if other error

values have been accepted for a pipette, then on selecting option

the error values are neglected, and replaced by

those from ISO standard),

Acquire data on ambient conditions from THB module: YES/NO. For

option the software automatically, and in specified moment,

acquires and saves ambient conditions’ parameters from a connected THB

module. If option is selected, before starting and after completing

the pipettes calibration process operator should manually insert ambient

conditions parameters (i.e. temperature, humidity and pressure in the

weighing room) from external ambient conditions sensors.

Result control, Tare mode, Automatic footer printout, Printout

mode, Printout, means of using these functions is given in point 10.8.

“Additional parameters on weighing process”.

24.2. Pipettes calibration – quick access keys

Each of working modes features a set of default quick access keys which are

automatically displayed on mode activation. The set of keys can be modified

by selecting other quick access keys to the bottom bar of the display. Such

process requires appropriate operator’s access level.

24.3. Adding a pipette to the database

Database of pipettes contains list of pipettes’ names and other data including

tested volume and errors values for a specific volume. When adding a pipette

to the database, first specify its name, and then add other data on a pipette.

The software operates intuitively, and commands a user to follow displayed

information.
Adding a pipette to the Database can be carried out in menu of the database.

Procedure:

Enter submenu < Databases>, press <

Pipettes> field

Press <

Add > soft key, should a new pipette be added