Use of checkweighing thresholds, Additional settings of checkweighing mode – RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 61
- 61 -
17.1. Use of checkweighing thresholds
Use of checkweighing thresholds can be carried out by:
−
Selection of Product <
> for which Low and High thresholds have been
already set
−
Inserting the numeric value of thresholds <
>, in this case the
thresholds are not referred to any product from database
PROCEDURE 1 – Selection of product from Database of Products
•
Press <
Database of Products> quick access key
•
Using list of products select one to be weighed
•
Automatically, in the grey workspace the values of thresholds are
displayed
•
Under the measurement result there is bargraph displayed. Colour of the
bargraph corresponds to the current status of mass:
o
Yellow bargraph’s colour: mass value below Low threshold
o
Green bargraph’s colour: mass value within set value of thresholds
o
Red bargraph’s colour: mass above set value of High threshold
PROCEDURE 2 – Manual inserting checkweighing thresholds
•
Press <
Checkweighing thresholds> quick access key
•
Press Low Threshold key and determine its value
•
Accept set value by pressing <
> key
•
Press High Threshold key and determine its value
•
Accept set value by pressing <
> key
Caution:
The value of High threshold has to be bigger than the value of Low threshold.
17.2. Additional settings of checkweighing mode
The additional settings enable adjusting the working mode to user’s needs and
requirements. To access the setting follow below procedure:
Procedure:
1. Press grey coloured workspace,
2. The display indicates menu of: Settings, Keys, Information, Printout and
Profile,
3. Press
4. The display indicates functions related to checkweighing:
−
Result control
−
Tare mode
−
Automatic footer printout
−
Printout mode / Value release
−
Printout
Means of using the above functions are given in point 10.8. “Additional
parameters on weighing process”.