Cooperation with titrators – RADWAG MYA 5.3Y.F1 Microbalance User Manual
Page 32

- 32 -
- Mass above the set value of minimum
sample for a given tare range
Meaning of the pictograms of minimum sample mode:
Mass below the set value of minimum sample.
Mass above or equal to the value of minimum sample.
Mass below the set value of minimum sample. A pictogram with a
clock informs on approaching expiring of the validity period of the
minimum sample (it is displayed two weeks before the
determined expiry date).
Mass above the set value of minimum sample. A pictogram with a
clock informs on approaching expiring of the validity period of the
minimum sample (it is displayed two weeks before the
determined expiry date).
The validity term of the minimum sample has expired.
Changes are needed in the minimum sample settings.
Only authorized RADWAG personnel can change the minimum
sample settings.
Caution: If more than one reference tare value is programmed (with
corresponding minimum sample mass) then the indicated value automatically
moves to a range corresponding to the weight of the tarred container. At the
same time the required minimum mass is changed.
10.9. Cooperation with TITRATORS
In order to ensure correct cooperation with TITRATORS it is necessary to set titrator> option to other variables for this printout must be off.