SERVICE

8

95

ALTOSONIC V12

www.krohne.com

04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en

• Loosen all coax cables on the bottom of the electronic unit (the number of cables depend on

the version of the flowmeter).

• Slide the new electronics unit partially back into the housing.
• Push the metal pullers back to their original position.

Don't use excessive force, otherwise the connector at the backside can be damaged!

• Screw the electronics unit back to the housing.
• Re-install the display and make sure not to kink the display's flat ribbon cable.
• Replace cover and tighten by hand.
• Connect power.

CAUTION!
Please pay attention that the same amount of force is applied on both pullers, otherwise the
connector at the backside can be damaged.

Figure 8-36: Location of connectors

1 Position of connectors for transducer cables
2 Metal pullers

DANGER!
Electrostatic discharge (ESD) can damage electronic parts. Make sure to discharge yourself by
wearing a wrist strap. If no wrist strap is available, ground yourself by touching a metal surface
that is grounded.