8

SERVICE

88

ALTOSONIC V12

www.krohne.com

04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en

Step 6: Mount the retraction tool

• Fasten the lower part of the retraction tool (the ball valve) on the transducer adapter by means

of 4 bolts M 8 x 130 (grade 12.9) using the 6 mm hexagon screw driver. Check the O-ring
sealing, if damaged replace (size 40.87 x 3.53 mm).

• Tighten the M8 bolts with appropriate force using the 6 mm hexagon angled key.
• Make sure that the ball valve is in fully open position.

• Measure the position of the transducer adapter inside the transducer port (depth in the center,

touching the bronze nut) with the ruler.

• Memorize and note this value as "Depth".

INFORMATION!
When properly tightened there should still be some clearance between both plates of the
adapter.

Figure 8-24: Mount the ball valve of the retraction tool to the flowmeter

1 Transducer, fastening nut and transducer adapter inside the flowmeter.
2 Wedge shaped adapter plate.
3 Ball valve of the retraction tool
4 Use a 6 mm hexagon key to tighten the four M8 x 130 bolts.

INFORMATION!
In the figure above the square wedge shaped adapter plate is shown. This can also be the round
adapter plate, dependent from the type of flowmeter.

Figure 8-25: Measure depth

1 Bronze fixation nut.
2 Insert ruler up to bronze fixation nut.
3 Read the depth.