Modbus protocol description and set-up – KROHNE ALTOSONIC V12 EN User Manual
Page 140

10
MODBUS PROTOCOL DESCRIPTION AND SET-UP
140
ALTOSONIC V12
www.krohne.com
04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en
Alarm configuration
21
7521
7542-7543
(2 regs)
Rev_LowFlowCutoff
Low Flow cutoff
reverse
For reverse flow rates with smaller magnitude
than this cutoff value, totalisers stop counting
and outputs (such as the frequency output)
will be disabled; flow rate indication will still
be available (in m/s).
22
7522
7544-7545
(2 regs)
Low_Flowcut_-
Threshold
Low Flow cutoff
Threshold
Hysteresis preventing frequent on/off
switching of low flow cut-off (in %).
Relative
address
Absolute address
Name
Parameter /
Variable
(short description)
non-Modicon
Modicon-comp.
Relative
address
Absolute address
Name
Parameter /
Variable
(short description)
non-Modicon
Modicon-comp.
23
7523
7546-7547
(2 regs)
SOSDev
Allowed Deviation
SOS
per channel on SOS
average pipe
Defines alarm limit for deviation or measured
SOS of any acoustic path from average value
of all paths (in m/s) (all SoS values calculate
with mach-correction).
24
7524
7548-7549
(2 regs)
T_LimitHigh
Input Temperature
Proces: LimitHigh
Upper limit for process temperature
measurement (in ° C).
25
7525
7550-7551
(2 regs)
T_LimitLow
Input Temperature
Proces: LimitLow
Lower limit for process temperature
measurement (in ° C).