Modbus protocol description and set-up – KROHNE ALTOSONIC V12 EN User Manual

Page 140

background image

10

MODBUS PROTOCOL DESCRIPTION AND SET-UP

140

ALTOSONIC V12

www.krohne.com

04/2013 - 4002643502 - MA ALTOSONIC V12 R02 en

Alarm configuration

21

7521

7542-7543

(2 regs)

Rev_LowFlowCutoff

Low Flow cutoff

reverse

For reverse flow rates with smaller magnitude

than this cutoff value, totalisers stop counting

and outputs (such as the frequency output)

will be disabled; flow rate indication will still

be available (in m/s).

22

7522

7544-7545

(2 regs)

Low_Flowcut_-

Threshold

Low Flow cutoff

Threshold

Hysteresis preventing frequent on/off

switching of low flow cut-off (in %).

Relative
address

Absolute address

Name

Parameter /
Variable
(short description)

non-Modicon

Modicon-comp.

Relative
address

Absolute address

Name

Parameter /
Variable
(short description)

non-Modicon

Modicon-comp.

23

7523

7546-7547

(2 regs)

SOSDev

Allowed Deviation

SOS

per channel on SOS

average pipe

Defines alarm limit for deviation or measured

SOS of any acoustic path from average value

of all paths (in m/s) (all SoS values calculate

with mach-correction).

24

7524

7548-7549

(2 regs)

T_LimitHigh

Input Temperature

Proces: LimitHigh

Upper limit for process temperature

measurement (in ° C).

25

7525

7550-7551

(2 regs)

T_LimitLow

Input Temperature

Proces: LimitLow

Lower limit for process temperature

measurement (in ° C).