Setup items – Yokogawa Removable Chassis DX1000N User Manual
Page 79
2-12
IM 04L41B01-01E
• Status Relay Details
Press
MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to basic
setting mode), and select the
Menu tab > Status Relay > Status Relay Details.
Setup Items
• Relay Action > Fail Relay, Status Relay (Release numbers 4 and later)
On DXs without the /AS1 advanced security option
Setting
Description
Fail
Outputs a fail relay.
Status relay
Outputs a status relay. Even if you set both relays to “Status relay,” you
cannot make them output different statuses.
On DXs with the /AS1 advanced security option
Setting
Description
Fail
Outputs a fail relay.
Status relay
Outputs a status relay. Even if you set both relays to “Status relay,” you
cannot make them output different statuses.
Mem. sample*
Relay output continues from memory start until memory stop.
Invalid user
Relay output continues from when a user is determined to be invalid until a
Locked ACK command is executed.
Login
Relay output occurs whenever a user is logged in.
* When the multi batch function (/BT2 option) is being used, relay output starts when memory
start occurs for any of the batches, and relay output stops when memory stop occurs for all
the batches.
• Status Relay Details > Memory/Media status
On: The status of the internal memory and the CF card is relayed.
• Status Relay Details > Measurement error
On: A relay signal is output when a measurement error occurs.
• Status Relay Details > Communication error
On: A relay signal is output when a communication error occurs.
• Status Relay Details > Memory stop
On: A relay signal is output when memory sampling stops.
* When the multi batch function (/BT2 option) is being used, relay output starts when
recording stops for all the batches.
• Status Relay Details > Alarm (Release numbers 3 and later)
On: Outputs a relay signal when one or more alarms occur. Alarms whose indications
are hidden (see section 3.6 for details) do not cause relay signal output. Relay output
ceases after all alarms are released (if no other relay output conditions are met).
2.9 Outputting the DX Status via the Relay Contact (/F1 and /F2 Options)