7.3 Changing the Response to Remote Contact
Input Opening and Closing (/R1 and /PM1
options; release number 3 or later)
Set the remote contact input operation.
Setup Screen
Press
MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to basic
setting mode), and select the
Environment tab > Remote.
Setup Items
• Remote Input > 1 to 8
Specify an operation for each remote control terminal. Five remote control terminals
can be used for pulse input (/PM1 option).
Setting
Description
N.O
The remote signal rises when the contact input switches from open to
closed, and it falls when the contact input switches from closed to open.
N.C
The remote signal rises when the contact input switches from closed to
open, and it falls when the contact input switches from open to closed.
Type
Remote
contact
input
Operation
Control operation
N.C
N.O
Closed
Closed
Closed
OFF
ON
Open
Open
Open
Note
On models with the pulse input option (/PM1), the remote control input terminals can be
used for pulse input. When pulses are counted, the number of rising pulse edges is counted,
regardless of the Remote Input setting.