USB Interface (/USB1)

Item

Specifications

USB port

Complies with Rev. 1.1 and host function

Number of ports

2 (one each on the front panel and rear panel)

Power supply

5 V , 500 mA (per port)

Devices which need more than 500 mA total bus power for 2 ports can not be connected at the

same time.

Connectable devices

Only connect the devices listed below to prevent damage to the devices.

Keyboard

Complies with HID Class Ver. 1.1

104 keyboard/89 keyboard (US) and 109 keyboard/89 keyboard (Japanese)

Number connectable units: 1

External medium

USB flash memory (up to 2 GB)

Does not guarantee the operation of all USB flash memories.

External medium such as a hard disk, ZIP, MO, and optical discs are not supported.

Number connectable units: 1

Barcode reader (release number 3 or later)

USB HID Class Ver. 1.1 compatible

English (U.S.) standard USB keyboard compatible

Pulse Input (/PM1)

Item

Specifications

Pulse input
Number of inputs

3 (8 when using the remote control input terminals)

Input type

Isolated from the main circuitry through a photocoupler and built-in isolated power supply for the

input terminals.

Shared common for pulse inputs.

Input type and signal level Voltage-free contact Contact closed at 200 Ω or less and contact open at 100 kΩ or greater

Open collector

ON voltage: 0.5 V or less (sink current 30 mA or more), leakage current when

OFF: 0.25 mA or less

Counting

Counts the rising edges of pulses.

For voltage-free contact input: Contact open to contact close

For open collector:

Voltage level of the terminal H from high to low

Allowable input voltage 30 VDC
Max. sampling pulse period 100 Hz
Minimum detected pulse width

5 ms or more for both low (closed) and high (open)

Pulse detection period

Approx. 3.9 ms (256 Hz)

Pulse measuring accuracy ±1 pulse
Pulse count interval

Scan interval or 1 s

Miscellaneous

Pulse input terminals can be used as remote control input terminals, isolated from remote control

input terminals

Remote control

Number of inputs: 5. Same as remote control (/R1) for the other specifications

Computation function

Same as the computation function (/M1)

Calibration Correction (/CC1)

Item

Specifications

Calibration correction method

Corrects the measured value of each channel using segment linearizer approximation.

Number of segment points: 2 to 16 (including the start and end points)

Calibration management

A function to make sure that calibration is performed regularly (release numbers 4 and later)

12.5 Options