Chapter 12 specifications, 1 signal input and alarm, 1 signal input and alarm -1 – Yokogawa Removable Chassis DX1000N User Manual
Page 295: App index, Measurement input
12-1
IM 04L41B01-01E
Specifications
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
12.1 Signal Input and Alarm
Measurement Input
Item
Specifications
Number of inputs, scan interval, and A/D integration time
DX1002, DX1002N
DX1004, DX1004N
DX1006, DX1006N
DX1012, DX1012N
25 ms
2
4
6
12
Fast Sampling Mode
*
Normal Mode
Number
of
Inputs
Model
Scan Interval
A/D converter integration time
60 Hz/50 Hz/100 ms
60 Hz/50 Hz
1 s, 2 s, 5 s
600 Hz (fixed)
125 ms
125 ms, 250 ms
2 s, 5 s
* When using the multi batch function (/BT2) the DX do not have a fast sampling mode.
Input Type
DC voltage, 1-5V, thermocouple (TC), resistance temperature detector (RTD), ON/OFF input (DI),
and DC current (by adding an external shut resistor)
Measurement range and measurable range
DC voltage
20 mV
–20.000 to 20.000 mV
60 mV
–60.00 to 60.00 mV
200 mV
–200.00 to 200.00 mV
2 V
–2.0000 to 2.0000 V
6 V
–6.000 to 6.000 V
20 V
–20.000 to 20.000 V
50 V
–50.00 to 50.00 V
1-5V
1 to 5 V*
6
0.800 to 5.200 V
Thermocouple R*
1
0.0 to 1760.0°C
S*
1
0.0 to 1760.0°C
B*
1
0.0 to 1820.0°C
K*
1
–200.0 to 1370.0°C
E*
1
–200.0 to 800.0°C
J*
1
–200.0 to 1100.0°C
T*
1
–200.0 to 400.0°C
N*
1
–270.0 to 1300.0°C, on models
with release numbers 3 and later.
0.0 to 1300.0°C, on models with
release numbers 2 and earlier.
W*
2
0.0 to 2315.0°C
L*
3
–200.0 to 900.0°C
U*
3
–200.0 to 400.0°C
WRe*
4
0.0 to 2400.0°C
RTD
Pt (Pt100)*
5
–200.0 to 600.0°C
JPt (JPt100)*
5
–200.0 to 550.0°C
DI
Level
0: Less than 2.4 V. 1: 2.4 V or higher (judged at the 6 V range)
Contact*
7
0: Open. 1: Closed (parallel capacitance of 0.01 µF or less)
*1: R, S, B, K, E, J, T, N: IEC584-1 (1995), DIN IEC584, JIS C1602-1995
*2: W: W-5%Re/W-26%Re (Hoskins Mfg. Co.), ASTM E988
*3: L: Fe-CuNi, DIN43710, U: Cu-CuNi, DIN43710
*4: WRe: W-3%Re/W-25%Re (Hoskins Mfg. Co.)
*5: Pt100: JIS C1604-1997, IEC751-1995, DIN IEC751-1996
JPt100: JIS C1604-1989, JIS C1606-1989
Measuring current: i = 1mA (Pt100, JPt100)
*6: The range for linear scaling of 1-5V inputs. Burnout detection and low-cut
functions are available.
*7: The detected current value is approx. 10 μA.
32 to 3200°F
32 to 3200°F
32 to 3308°F
–328 to 2498°F
–328.0 to 1472.0°F
–328.0 to 2012.0°F
–328.0 to 752.0°F
–454 to 2372°F
, on models with
release numbers 3 and later.
32 to 2372°F
, on models with
release numbers 2 and earlier.
32 to 4199°F
–328.0 to 1652.0°F
–328.0 to 752.0°F
32 to 4352°F
–328.0 to 1112.0°F
–328.0 to 1022.0°F
Measurable Range
Range
Input Type
Chapter 12 Specifications