beautypg.com

Chapter 12 specifications, 1 signal input and alarm, 1 signal input and alarm -1 – Yokogawa Removable Chassis DX1000N User Manual

Page 295: App index, Measurement input

background image

12-1

IM 04L41B01-01E

Specifications

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

12.1 Signal Input and Alarm

Measurement Input

Item

Specifications

Number of inputs, scan interval, and A/D integration time

DX1002, DX1002N
DX1004, DX1004N
DX1006, DX1006N
DX1012, DX1012N

25 ms

2
4
6

12

Fast Sampling Mode

*

Normal Mode

Number

of

Inputs

Model

Scan Interval

A/D converter integration time

60 Hz/50 Hz/100 ms

60 Hz/50 Hz

1 s, 2 s, 5 s

600 Hz (fixed)

125 ms

125 ms, 250 ms

2 s, 5 s

* When using the multi batch function (/BT2) the DX do not have a fast sampling mode.

Input Type

DC voltage, 1-5V, thermocouple (TC), resistance temperature detector (RTD), ON/OFF input (DI),

and DC current (by adding an external shut resistor)

Measurement range and measurable range

DC voltage

20 mV

–20.000 to 20.000 mV

60 mV

–60.00 to 60.00 mV

200 mV

–200.00 to 200.00 mV

2 V

–2.0000 to 2.0000 V

6 V

–6.000 to 6.000 V

20 V

–20.000 to 20.000 V

50 V

–50.00 to 50.00 V

1-5V

1 to 5 V*

6

0.800 to 5.200 V

Thermocouple R*

1

0.0 to 1760.0°C

S*

1

0.0 to 1760.0°C

B*

1

0.0 to 1820.0°C

K*

1

–200.0 to 1370.0°C

E*

1

–200.0 to 800.0°C

J*

1

–200.0 to 1100.0°C

T*

1

–200.0 to 400.0°C

N*

1

–270.0 to 1300.0°C, on models

with release numbers 3 and later.

0.0 to 1300.0°C, on models with

release numbers 2 and earlier.

W*

2

0.0 to 2315.0°C

L*

3

–200.0 to 900.0°C

U*

3

–200.0 to 400.0°C

WRe*

4

0.0 to 2400.0°C

RTD

Pt (Pt100)*

5

–200.0 to 600.0°C

JPt (JPt100)*

5

–200.0 to 550.0°C

DI

Level

0: Less than 2.4 V. 1: 2.4 V or higher (judged at the 6 V range)

Contact*

7

0: Open. 1: Closed (parallel capacitance of 0.01 µF or less)

*1: R, S, B, K, E, J, T, N: IEC584-1 (1995), DIN IEC584, JIS C1602-1995

*2: W: W-5%Re/W-26%Re (Hoskins Mfg. Co.), ASTM E988

*3: L: Fe-CuNi, DIN43710, U: Cu-CuNi, DIN43710

*4: WRe: W-3%Re/W-25%Re (Hoskins Mfg. Co.)

*5: Pt100: JIS C1604-1997, IEC751-1995, DIN IEC751-1996

JPt100: JIS C1604-1989, JIS C1606-1989

Measuring current: i = 1mA (Pt100, JPt100)

*6: The range for linear scaling of 1-5V inputs. Burnout detection and low-cut

functions are available.

*7: The detected current value is approx. 10 μA.

32 to 3200°F

32 to 3200°F

32 to 3308°F

–328 to 2498°F

–328.0 to 1472.0°F

–328.0 to 2012.0°F

–328.0 to 752.0°F

–454 to 2372°F

, on models with

release numbers 3 and later.

32 to 2372°F

, on models with

release numbers 2 and earlier.

32 to 4199°F

–328.0 to 1652.0°F

–328.0 to 752.0°F

32 to 4352°F

–328.0 to 1112.0°F

–328.0 to 1022.0°F

Measurable Range

Range

Input Type

Chapter 12 Specifications

This manual is related to the following products: