Transport and storage conditions, Supported standards – Yokogawa Removable Chassis DX1000N User Manual
Page 314
12-20
IM 04L41B01-01E
Transport and Storage Conditions
Item
Specifications
Ambient temperature
–25 to 60°C
Ambient humidity
5 to 95%RH (no condensation)
Vibration
10 to 60 Hz, 4.9 m/s
2
maximum
Shock
392 m/s
2
maximum (in packaged condition)
Supported Standards
Item
Specifications
CSA
CSA22.2 No.61010.1, installation category II
*1
, pollution degree 2
*2
, and measurement category II
*3
UL
UL61010-1 (CSA NRTL/C)
CE
EMC directive
EN61326-1 compliance, Class A Table 2
EN61000-3-2 compliance
EN61000-3-3 compliance
EN55011 Class A Group 1
Low voltage directive
EN61010-1, installation category II
*1
, pollution degree 2
*2
, and measurement category II
*3
C-Tick
EN55011 compliance, Class A Group 1
*1 Installation category (overvoltage category) II: Describes a number which defines a transient overvoltage condition. Iimplies the regulation for
impulse withstand voltage. “II” applies to electrical equipment which is supplied from the fixed installation like a distribution board.
*2 Pollution degree 2: Describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface resistivity is adhering. “2”
applies to normal indoor atmosphere. Normally, only non-conductive pollution occurs.
*3 Measurement category II: Applies to measuring circuits connected to low voltage installation, and electrical instruments supplied with power
from fixed equipment such as electric switchboards.
12.6 General Specifications