3.12 Using the Alarm Annunciator Function (Release

number 3 or later)

This section explains how to use the alarm annunciator function.
For a description of the function, see section 1.2.

Setup Screen

• Annunciator Mode

Press

MENU (to switch to setting mode), hold down FUNC for 3 s (to switch to basic

setting mode), and select the

Environment tab > Alarm > Action.

• Alarm Levels and Colors

See section 3.7.

• Display

Press

MENU (to switch to setting mode), and select the Menu tab > Display >

Annunciator.

Setup Items

• Alarm action > Annunciator mode

Select

On to use the annunciator function.

• Alarm action > Sequence

Select the annunciator sequence. For details about the annunciator sequence, see

“Explanation” in this section.

Setting

Description

ISA-A-4

A no lock-in sequence.

ISA-A

A lock-in sequence.

ISA-M

A double lock-in sequence.