Warning Messages
Code
Message
Ref. Section
600
Measured data and Settings have been initialized.
–
601
Measured data have been initialized.
–
610
This usernema is already registered.
See section 8.2.
611
There is no user who can enter to the SETUP mode.
–
612
Please acknowledge all active alarms before stopping this record.
This message appears if you try to stop
recording when there is an alarm that the alarm
ACK operation has not been performed on. You
can either stop recording or perform the alarm
ACK operation and then stop recording. See
the See sections 6.3 and 6.4 in the advanced
security manual.
614
Calibration settings are reset because of range setting change.
See section 3.9.
615
Noise may influence measurement in test mode.
See section 3.1.
For information about messages whose codes are in the 700s, see page 11-5.
System Errors
Servicing is required when a system error occurs. If this happens, contact your nearest
YOKOGAWA dealer for repairs.
Code
Message
901
ROM failure.
902
RAM failure.
910
A/D memory failure for all input channels.
921
Channel 1 A/D calibration value error.
925
A/D calibration error.
930
Memory acquisition failure.
940
The Ethernet module is down.
950
Incorrect number for the A/D calibration.
951
Failed to write A/D calibration value.
10.1 A List of Messages