6-13

IM 04L41B01-01E

Saving and Loading Data

1

2

3

4

5

6

7

8

9

10

11

12

App

Index

• Changing Settings and Performing File Operations during Recording

(Memory sampling)

If you are using the multi batch function (/BT2 option), if even one batch group is being

recorded, the DX is recording (memory sampling).

On DXs without the /AS1 advanced security option

You can change all settings except for the ones listed below.

• Basic settings
• Input range
• Memory sampling on/off
• Computation channel calculation expressions and constants
• TLOG
• Trend interval
• File save interval
• Timer and match time timer

On DXs With the /AS1 Advanced Security Option

You can change the following settings and perform the following file operations. The

administrator can perform all operations. Users can only perform operations that have
been permitted. The setting menu that appears varies depending on the operations
that can be performed.

Setting Changes

• Date and time settings
• Alarm settings
• Alarm delay time settings
• Calibration settings
• Destination directory settings
• Administrator settings*
• User settings*

* See the Advanced Security Function (/AS1) User’s Manual, IM 04L41B01-05EN.

File Operations

• Loading display data files
• Loading event data files
• Listing the files on the external storage medium

6.4 Starting/Stopping the Recording and Saving the Measured data