2 alarms, 2 alarms -4, Alarm type – Yokogawa Removable Chassis DX1000N User Manual
Page 15
1-4
IM 04L41B01-01E
1.2 Alarms
This function generates an alarm when the measured data meets a certain condition. Up
to four alarms can be set for each channel.
Alarm Type
You can use the alarms shown below. The character inside the parentheses is the
symbol denoting each alarm.
For the alarm setting procedure, see section 3.7.
High Limit Alarm (H)
Alarm value
Alarm output ON
Alarm output ON
Alarm output ON
Alarm output ON
Alarm value
Hysteresis
Measured value
Measured value
Hysteresis
Low Limit Alarm (L)
Alarm output ON
Alarm output ON
Delay time
Delay time
Alarm value
Alarm value
Measured value
Measured value
Delay High Limit Alarm (T)
Delay Low Limit Alarm (t)
High Limit on Rate-of-Change Alarm (R)
Low Limit on Rate-of-Change Alarm (r)
Alarm value
Alarm value
Hysteresis
Hysteresis
Difference in the measured values of two channels
Difference in the measured values of two channels
Difference High Limit Alarm (h)
Difference Low Limit Alarm (l)
Change in the
measured value
Change in the measured value
Measured value
Measured value
Time
Time
Interval (t2-t1)
Interval (t2-t1)
Amount of change in
the setting
|
|
T2
T1
t1
t2
T2
T1
t1
t2
T2-T1
Amount of change in
the setting
|
T2-T1
• Alarm Hysteresis
You can set a width (hysteresis) to the value used to activate and release alarms.
• Delay High Limit Alarm and Delay Low Limit Alarm
An alarm occurs when the measured value remains above or below the alarm value
for a specified time period (delay period).
• High Limit on Rate-of-Change Alarm and Low Limit on Rate-of-Change
Alarm
The rate-of-change of the measured values is checked over a certain time (interval).
An alarm occurs if the rate-of-change of the measured value in the rising/falling
direction is greater than or equal to the specified value.
The alarm value of the rate-of-change alarm is set using an absolute value. The
interval is derived using the following equation and set using the number of samples.
Interval = the scan interval × the number of samples
For the setting procedure, see section 3.5.
• Difference Upper Limit Alarm and Difference Lower Limit Alarm
An alarm occurs when the difference in the measured values of two channels is
greater/less than or equal to the specified value. These alarms can be specified on
measurement channels set to difference computation.