Special data handling – Yokogawa Removable Chassis DX1000N User Manual
Page 61
1-50
IM 04L41B01-01E
Special Data Handling
This section explains the handling of special data in TLOG computation, CLOG
computation, and reports.
• Power Failure Operation (TLOG and Reports)
If a power failure occurs when the report function is enabled or in the middle of the
TLOG computation, the report operation and TLOG computation resume when the
DX recovers from the power failure. The operation varies depending on whether the
DX recovers from the power failure before or after the time to create a report or TLOG
data.
Time of Recovery
Operation
After the time to create the data
The report or TLOG data is created immediately after the DX
recovers. The measured data up to the time of the power
failure is used. At the time the next report or TLOG data is
created, the data after the recovery is used.
Before the time to create the data After the DX recovers, report or TLOG data is created at the
normal time to create the data. The measured data excluding
the power failure period is used.
• Error Data Handling (TLOG, CLOG, and Reports)
If an error occurs in the channel data, the error data is discarded, and the computation
continues. If all the data are in error, an error results.
The following types of data are considered error data.
• Channels set to skip or Off.
• The measured result on a measurement channels is error (A/D converter failure,
etc.).
• The input of the measurement channel is in a burnout condition.
• The computed result on a computation channel is error.
• Handling of Overflow Data*
* Refers to over range on a measurement channel and computation overflow on a computation
channel.
For TLOG, CLOG, and Reports
When the channel data is overflow data, the DX handles the data as follows:
Computation Type
Description
Average value or sum Set the handling to ERROR, SKIP, or LIMIT.
value
ERROR: Considers the data to be a computation error.
SKIP:
Discards the overflow data and continues the computation.
LIMIT:
Replaces the data with the limit value and continues the
computation.
The limit value is the span upper or lower limit or the scale
upper or lower limit of the channel.
Maximum, minimum,
Set the handling to OVER or SKIP.
Maximum – minimum OVER: Computes by using the overflow data.
SKIP:
Discards the overflow data and continues the computation.
For Multiplication and Relation Computation EQ and NE
Computation Type
Computation
Computed Result
Multiplication
0*(+Over)
0
0*(–Over)
0
(+Over)*0
0
(–Over)*0
0
.EQ.
(+Over).EQ.(+Over)
0
(–Over).EQ.(–Over)
0
.NE.
(+Over).NE.(+Over)
1
(–Over).NE.(–Over)
1
1.8 Computation and Report Function (/M1 and /PM1 Options)