Status Output

Outputs the status below with a relay contact signal (1 relay). You can select whether
each status is output to the relay. The relay is energized when the status occurs. You
cannot change this behavior.
For the setting procedure, see section 2.9.

Status

Description

Corrective Action

Status of the internal
memory or CF card

Error in the internal memory.

Contact your nearest

YOKOGAWA dealer for

repairs.

When the auto save function to the CF card is On.

• The free space on the CF card dropped

to 10% of the total size (only when the

media FIFO (see section 1.4) is disabled).

Replace the CF card.

• Error in the CF card.

• Replace the CF card with a

normal one.

• Format the CF card on the

DX (the data on the CF

card will be erased).

However, the status of the internal memory

is output when the CF card is not inserted.
• 10 MB or less of available space*

remaining in internal memory.

• The number of files in internal memory for

which Auto Save to the CF card has not

been completed has exceeded 390.

Insert a CF card.

When the auto save function to the CF card is Off.

• 10 MB or less of available space*

remaining in internal memory.

• The number of files in internal memory

for which Manual Save has not been

completed has exceeded 390.

Save the data in the internal

memory to the CF card.

Measurement error

Error in the A/D converter.
Burnout is detected.

Contact your nearest

YOKOGAWA dealer for

repairs.
Replace the thermocouple

that has burned out.

Communication error A Modbus master or Modbus client

communication error occurred.

Check the error in the Modbus

master or Modbus client

screen and carry out the

corrective action.

A PROFIBUS-DP communication error

occurred.

Contact your nearest

YOKOGAWA dealer for

repairs.

Memory stop

When the memory sampling is stopped.

Start the data acquisition.

Alarm occurrence

(Release number 3 or

later)

An alarm has occurred.

Check the alarm.

* The internal memory’s “available space” refers to the following quantities.
• Unused regions
• Regions of data for which Auto Save or Manual Save (see page 1-31) has been

completed.

Relay Operation

NO

C

NC

During normal operation

NO

C

NC

When specified status occurs

NO

C

NC

When power is turned OFF

1.9 FAIL/Status Output Function (/F1 and /F2 Options)