1.5 Specifications

Package

Dimensions:

10.5 x 10.4 x 2.4 inches
(26.7 x 26.5 x 6 centimeters)

Connectors:

one USB (type “B”)
three 1394 (6-pin)

Weight:

3.5 pounds
(1.6 kilograms)

Power Requirements

90-264VAC, 47-63Hz (universal input), 100W maximum

Environmental Conditions

Operating Range: 0 to 55 °C (32 to 131 °F)
Storage Range:

-20 to 80 °C (-4 to 176 °F)

Humidity:

10 to 90%, non-condensing

LEDs

Power (PWR):

illuminates when the analyzer is powered on

Triggered (TRG): illuminates when the analyzer has detected a valid trigger condition

Recording (REC): illuminates when the analyzer is actively recording traffic data

Recording Memory Size

64 MB DRAM for traffic data capture
64 MB DRAM for timing, state & other data

Certification

FCC (Class A), CE Mark, UL, CSA

Basic Events Detected

FireInpector detects the following basic events:

• Bus conditions: bus reset, arbitration reset gap, subaction gap, and packet speed

• PHY packets

22

GND

Signal Ground

23

GND

Signal Ground

Table 1-2: External Interface Breakout Board – Pin Description

Pin

Signal Description