beautypg.com

Appendix a, Certifications, Emc compliance – Teledyne LeCroy PeRT3 Phoenix System User Manual

Page 113

background image

Teledyne LeCroy PeRT

3

Phoenix System User Manual

1111
11

Appendix A






Certifications

This section certifies the instrument's Electromagnetic Compatibility (EMC), Safety and
Environmental compliances.

EMC Compliance

EC DECLARATION OF CONFORMITY - EMC

The instrument meets intent of EC Directive 2004/108/EC for Electromagnetic
Compatibility. Compliance was demonstrated to the following specifications as listed in
the Official Journal of the European Communities:

EN 61326‐1:2006, EN 61326‐2‐1:2006 EMC requirements for electrical equipment for
measurement, control, and laboratory use. 1

Electromagnetic Emissions:

CISPR 11:2003, Radiated and Conducted Emissions Group 1, Class A 2 3

EN 61000‐3‐2:2006 Harmonic Current Emissions, Class A

EN 61000‐3‐3/A2:2005 Voltage Fluctuations and Flickers, Pst = 1

Electromagnetic Immunity:

EN 61000‐4‐2:2001 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV vertical/horizontal
coupling planes 4

EN 61000‐4‐3:2006 RF Radiated Electromagnetic Field, 3 V/m, 80‐1000 MHz; 3 V/m, 1400
MHz ‐ 2 GHz; 1 V/m, 2 GHz ‐ 2.7 GHz 4

EN 61000‐4‐4:2004 Electrical Fast Transient/Burst, 1 kV on power supply lines, 0.5 kV on I/
O signal data and control lines 4

EN 61000‐4‐5:2006 Power line Surge, 1 kV AC Mains, L‐N, L‐PE, N‐PE 4

EN 61000‐4‐6:2007 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz ‐ 80 MHz 4

EN 61000‐4‐11:2004 Mains Dips and Interruptions, 0%/1 cycle, 70%/25 cycles, 0%/250
cycles 4 5

1. To ensure compliance with all applicable EMC standards, high quality shielded

interface cables should be used.