Calibration Teledyne LeCroy

Calibration

The instrument is calibrated at the factory prior to being shipped. The recommended
calibration interval is one year. Calibration should be performed by qualified personnel
only.

The instrument software includes automatic and manual calibration functions.

CAUTION. It is required that all inputs be removed from the instrument prior to

performing a manual calibration.

Schedule an annual factory calibration as part of your regular maintenance. Extended
warranty, calibration, and upgrade plans are available for purchase. Contact your
Teledyne LeCroy sales representative or [email protected] to
purchase a service plan.

Power

Power Consumption

Power Consumption (Nominal): 325 Watts (325 VA)

Power and Ground Connections

The instrument is provided with a 10A/250V 18AWG rated grounded cord set containing a
molded three‐terminal polarized plug and a standard IEC320 (Type C13) connector for
making line voltage and safety ground connections.

The AC inlet ground is connected directly to the frame of the instrument. For adequate
protection again electric shock, connect to a mating outlet with a safety ground contact.

WARNING. Interrupting the protective conductor inside or outside the

instrument, or disconnecting the safety ground terminal, creates a hazardous
situation. Intentional interruption is prohibited.