Runt Trigger

The Runt trigger is programmed to occur when a pulse crosses a first threshold line and fails to cross a second
threshold line before recrossing the first. You can select both voltage thresholds within a time range of 100 ps to
20 s. Other defining conditions for this trigger are the edge (triggers on the slope opposite to that selected) and
runt width.

R

UNT

T

RIGGER

S

ETUP

1. Touch

the

Trigger descriptor label.

2. If applicable, touch the

Smart trigger button, then touch the Runt trigger button.

3. Touch inside the trigger

Source field and select a source on which to trigger. If

4. you select an external trigger source, you will also need to set a coupling mode.

5. Touch

inside

the

Level data entry field and enter a value, using the pop-up keypad.

6. Select Positive or

Negative polarity.

7. Touch

the

Less Than button and enter a pulse-width value in the Upper Limit field.

o

Or touch the

Greater Than button and enter a pulse-width value in the Lower Limit field.

o

Or touch the

In Range button. Then touch the Delta button to set up a nominal range, plus or minus a

delta value in seconds. Touch inside the

Nominal Width and Delta fields and enter values, using the

pop-up numeric keypads. Alternatively, touch the

Limits button to set up a precise pulse-width range

by setting

Lower Limit and Upper Limit values, using the pop-up keypads.

o

Or touch the

Out Of Range button and perform the same range setups as for In Range triggering.

Slew Rate Trigger

1. Touch

the

Trigger descriptor label.

2. If applicable, touch the

Smart trigger button, then touch the Slew Rate trigger button.

3. Touch inside the trigger

Source field and select a source on which to trigger. If

4. you select an external trigger source, you will also need to set a coupling mode.

5. Touch

inside

the

Level data entry field and enter a value, using the pop-up keypad.

6. Select

Positive or Negative polarity.

7. Touch

the

Less Than button and enter a pulse-width value in the Upper Limit field.

o

Or touch the

Greater Than button and enter a pulse-width value in the Lower Limit field.

o

Or touch the

In Range button. Then touch the Delta button to set up a nominal range, plus or minus a

delta value in seconds. Touch inside the

Nominal Width and Delta fields and enter values, using the

pop-up numeric keypads. Alternatively, touch the

Limits button to set up a precise pulse-width range

by setting

Lower Limit and Upper Limit values, using the pop-up keypads.

o

Or touch the

Out Of Range button and perform the same range setups as for In Range triggering.

DISPLAY FORMATS

1. In the menu bar, touch

Display; then touch Display Setup in the drop-down menu.

WRXi-OM-E Rev C

71

2. Touch one of the

Grid combination buttons. Autogrid automatically adds or deletes grids as you select more

or fewer waveforms to display.