6. Touch

inside

the

Condition field in the ParamCompare mini-dialog and select a math operator from the pop

up menu:

-

.

7. Touch

inside

the

Limit field and enter a value, using the pop-up numeric keypad. This value takes the

dimensions of the parameter that you are testing. For example, if you are testing a time parameter, the unit is

WRXi-OM-E Rev C

145

seconds. If you chose either

WithinDeltaPct

or

WithinDeltaAbs

from the

Condition menu,

you also have the choice of setting the limit by means of the statistical buttons at the bottom of the
ParamCompare dialog:

Comparing Dual Parameters

1. Touch

Analysis in the menu bar, then Pass/Fail Setup... in the drop-down menu.

2. Touch

a

Qx tab; a setup dialog for that position

inside

the

Condition field in the main dialog and select DualParamCompare

will open.

3. Touch

4. Touch

inside

the

Source1 and Source2 fields and select a source from the pop-up menu.

ompare mini-dialog field and select a source from the pop-up menu.

Compare Values field and select All or Any from the pop-up menu:

5. Touch

inside

the

ParamC

6. Touch

inside

the

By selecting

, the test will be true only if every waveform falls

All

within the limit that you will set. By selecting

Any, the test will be

st one waveform falls w

it.

true if ju

ithin the lim