Statistics, Applying a measure mode, Measure modes – Teledyne LeCroy WaveRunner XI SERIES Operator’s Manual User Manual

Page 102

background image

W

AVE

R

UNNER

X

I

S

ERIES

102

WRXi-OM-E Rev C

Parameter in error (P1) Out Result

4. Select Measure

5. Read the status information in line

StatusDescription.

Statistics

By touching the

Statistics On checkbox in the Measure, you can display statistics for standard vertical or

horizontal parameters, or for custom parameters. The statistics that are displayed are as follows:

value (last)

mean

min.

max.

sdev

num

The values displayed in the

num row are the number of measurements computed. For

computes on an entire waveform (like edge@level, mean, minimum, maximum, etc.) th

any parameter that

e value displayed

, the value displayed is equal to the number of events per

1. In the menu bar, touch

Measure, then Measure Setup.

2. Choose

a

Measure Mode from the dialog. The parameters are displayed below the grid.

Measure Modes

The selections for Measure Mode allow you to quickly apply parameters for standard vertical and standard
horizontal setups, and custom setups.

represents the number of sweeps.

For any parameter that computes on every event
acquired waveform. If

x waveforms were acquired, the value represents x times the number of cycles per

waveform. Also, the "value" is equal to the measurement of the last cycle on the last acquisition.

Applying a Measure Mode