4. If applicable, touch the

Pattern tab. For each input you want to include in the logic pattern, touch inside the

State field and select a logic state: Low or High. Select Don't Care for all other inputs.

5. Touch

inside

the

Level data entry field for each input included in the pattern, and enter a voltage level

threshold, using the pop-up keypad.

TV Trigger

TV triggers provide stable triggering on standard or custom composite video signals. Use them on PAL, SECAM,
or NTSC systems.

TV

T

RIGGER

S

ETUP

1. Touch

the

Trigger descriptor label.

2. In the Trigger dialog, touch the

TV trigger button

3. Touch one of these buttons (NTSC or PAL/SECAM) to select a television

standard.

4. Touch

inside

TV Trigger Source and select the Ext (location of Trigger Out

cable) from the pop-up menu.

Note: The channel that you have the VT75 Module connected to is
automatically designated TV Signal in the menu.

5. Touch

inside

Polarity and select positive or negative polarity. This selection is

independent of the polarity selected for Edge trigger (positive) in the "Trigger"
setup dialog.

6. Check

the

Trigger on Any Line checkbox if you do not want to specify one. In

this case the oscilloscope will trigger on any sync pulse.

7. Touch

inside

# of Fields and make a selection (1 to 8) from the pop-up menu.

8. Touch

inside

Line and enter a line number, using the pop-up keypad. The

maximum line number that you can enter depends on the television standard
(NTSC/PAL/SECAM) and the interlace factor selected. For NTSC, line 526 is
the same as line 1; for PAL, line 626 is the same as line 1.

9. Touch

inside

Field and make a selection from the pop-up menu. The maximum

field number that you can select is limited by your selection in

# of Fields. If # of

Fields above is set to

1, this selection is disabled.