beautypg.com

Warranty and service, Environmental characteristics, Temperature – Teledyne LeCroy WaveRunner XI SERIES Operator’s Manual User Manual

Page 21: Humidity, Altitude, Random vibration, Shock, Certifications, Ce declaration of conformity

background image

O

PERATOR

S

M

ANUAL

Power Consumption: 340 watts (340 VA) max., WaveRunner 62Xi: 290 W (290 VA), depending on

accessories installed (probes, PC port plug-ins, etc.); Standby State: 12 watts

Physical Dimensions (HWD): 260 mm x 340 mm x 152 mm (10.2 in. x 13.4 in. x 6.0 in.); height

measurement excludes foot pads

Weight: 6.95 kg (15.3 lbs.)

Warranty and Service

3-year warranty; calibration recommended yearly

Optional service programs include extended warranty, upgrades, and calibration services.

Environmental Characteristics

T

EMPERATURE

Operating: 5 to 40 °C
Storage (non-operating): -20 to +60 °C

H

UMIDITY

Operating: Maximum relative humidity 80% for temperatures up to 31 °C decreasing linearly to 50% relative

humidity at 40 °C.

Storage (non-operating): 5 to 95% RH (non-condensing) as tested per MIL-PRF-28800F

A

LTITUDE

Operating: Up to 2,000 m
Storage (non-operating): 12,192 m (40,000 ft)

R

ANDOM

V

IBRATION

Operating: 5 Hz to 500 Hz, overall level: 0.31 g

rms

, 15 minutes in each of 3 orthogonal axes

Non-operating: 5 Hz to 500 Hz, overall level: 2.4 g

rms

, 15 minutes in each of 3 orthogonal axes

S

HOCK

Functional Shock: 20 g peak, half sine, 11 ms pulse, 3 shocks (positive and negative) in each of 3

orthogonal axes, 18 shocks total

Certifications

CE Compliant, UL and cUL Listed

CE Declaration of Conformity

The oscilloscope meets requirements of EMC Directive 89/336/EEC for Electromagnetic Compatibility and Low
Voltage Directive 73/23/EEC for Product Safety.

EMC Directive:

• EN

61326/A3:2003

• EMC requirements for electrical equipment for measurement, control,

and laboratory use.

Electromagnetic Emissions:

• EN 55011/A2:2002, Radiated and conducted emissions (Class A)*
• EN 61000-3-2/A2:2005 Harmonic Current Emissions (Class A)
• EN 61000-3-3/A2:2005 Voltage Fluctuations and Flickers (Pst = 1)

* To conform to Radiated Emissions standard, use properly shielded cables on all I/O terminals.

WRXi-OM-E Rev C

21