beautypg.com

Example 4, Example 5 – Teledyne LeCroy WaveRunner XI SERIES Operator’s Manual User Manual

Page 170

background image

W

AVE

R

UNNER

X

I

S

ERIES

E

XAMPLE

4

Example 4 is a measurement of DVI (Digital Video Interface) Data-Clock skew jitter measurement, using a
VBScript to emulate the PLL.

In this example, a customer was not able to probe the desired clock signal. The only probing point available was

clock was a factor of 10 slower than the clock embedded in

the data signal (C3). By using a VBScript to create a clock waveform of the appropriate frequency (waveform F1),
the customer was able to display and measure data-clock skew using a LeCroy instrument function and
parameter.

E

XAMPLE

5

the output differential clock signal (C2). However, that

Next is a logarithmic vertical scale, for which the script can be found here. (Most scripts would be far simpler than
this one.)

Frequency response curves are frequently drawn on a logarithmic scale. The upper trace is a frequency sp
of a square wave after enhanced resolution has been applied. It was created using instrum

ectrum

ent functions. The

lower trace is the first lobe of the FFT display. But with a logarithmic frequency scale. Click here for the VBScript.

170