The alarm and warning limits for inputs can be changed in the configuration menu in the menu of the
desired input.
The alarm and warning limits for oxygen (module O2_SCAV_CTRL_RATIO) can be set in the
"Parameter" main menu in the submenu belonging to the module.
Configuration (Code Level: Specialist)
Configuration
Access
via Cfg.-
menu
Access
via XML-
Cfg.-File
Abbrevi
ation
(menu)
Value range
Default values
(after factory reset or
at start of Cfg-File-
Download)
Alarm output function
rw
rw
Output
enable/disable
disable
Output used for alarm
output
r rw Alarm
output
depending on
configured and
activated outputs and
adapted options:
Relay 1 … Relay 5,
4…20mA output 1 …
4…20mA output 4,
Transistor 1 …
Transistor 4,
Not output
No Output
The output used as alarm output has to be a configured and enabled output, which is not used
concurrently as output of another module.
The yellow LED indicates also in case of disabled alarm output function that an alarm occurs.
•