beautypg.com

Atec Agilent-8563E User Manual

Page 92

background image

ESG Family Signal Generators

Performance Tests

22. Dual Arbitrary Waveform Generator Check (Option UND only)

Calibration Guide

4-43

3. Verify that Channel 1 and Channel 2 are both displaying triangular waveforms

approximately 180 degrees out of phase and that they do not have any discontinuities.
Refer to the illustration below.

• Frequency is approximately 64 kHz.

• Amplitude is approximately 1.9 V

p-p

.

• Both waveforms are symmetrical around 0 volts (the average voltage is

approximately 0V).