beautypg.com

Custom i/q rf modulation quality (optionun8only), Recommended equipment, Equipment setups – Atec Agilent-8563E User Manual

Page 78: Custom i/q rf modulation quality (option un8 only)

background image

ESG Family Signal Generators

Performance Tests

14. Custom I/Q RF Modulation Quality (Option UN8 only)

Calibration Guide

4-29

14. Custom I/Q RF Modulation Quality
(Option UN8 only)

This automated test verifies the RF modulation quality of the Option UN8 signal
generator’s internal real time I/Q baseband modulation. A vector signal analyzer is
connected to the signal generator’s RF output. The Option UN8 internal baseband
generator modulates the RF carrier in each of the TDMA digital modulation formats. The
parameters of the individual TDMA formats common to Options UN3, UN4 and UN8 are
tested elsewhere. The vector signal analyzer measures the appropriate error parameter for
the modulation generated:

• EVM for QPSK,

π

/4DQPSK, and QAM16

• Phase error for MSK

• Frequency Shift Error (%) for FSK2

Recommended Equipment

• MD/MDC-174 Mixer (0.001 to 2.8 GHz)

• HP/Agilent 8663A Signal Generator

• HP/Agilent 89441A Options AY9, AYA, AYH, UFG or UTH Vector Signal Analyzer

Equipment Setups

Figure 4-26 Custom I/Q RF Modulation Quality

2000 MHz Test Setup