ESG Family Signal Generators

Service Software

ESG Family Support Software Administration

Calibration Guide

3-9

Figure 3-3

Adding Test Equipment Using the Test Equipment Menu

3. Select the model of the device.

4. Click Add.

5. In the New Test Equipment window (refer to

Figure 3-4

), enter the following

parameters into the appropriate box:

Serial Number

of the new equipment.

GPIB Address

of the new equipment. This address must be in the range of 0
through 31 and it should not conflict with any other instrument
address already present in the test setup.

NOTE

The power sensor must be assigned GPIB address

1 (negative 1).

Cal Due Date

the projected calibration due date of the new equipment.

Trace Number

the calibration tracking number. This is the last required item.

Calibration Type

(optional) is used only when special calibration data is associated

with the device being added, and only as a function setup by the
factory for equipment requiring specific calibration data. The only
devices currently requiring this feature are power sensors.

Table 3-1 is an example of the calibration information required

to ensure accuracy for measurements using the power sensor.
This window is accessed for data entry by selecting the words
CAL DATA from the Calibration Type field in the Edit or New
Test Equipment window,

Figure 3-4

.