Atec Agilent-8563E User Manual
Page 2
Equipment Required
ESG Family Signal Generators
Required Equipment
1-4
Calibration Guide
Low Noise Amplifiers for
Power Level Accuracy Adjustment and Performance Test
The power level accuracy adjustment and performance test both use low noise amplifiers.
They are required to maintain separation between a low-level signal and the signal
analyzer noise floor. The following amplifiers meet the requirements of making this
measurement.
A +15 Vdc external supply is required for each of the amplifiers. An HP/Agilent E3615A,
0 to 20 V power supply can power all recommended amplifiers simultaneously.
100 kHz to 10 MHz Low Noise Amplifier
Critical Specifications
Frequency Range: 100 kHz to 10 MHz
Noise Figure: 1.5 dB
Gain: 50 dB
1 dB Compression Minimum: +5 dBm
External Power Supply Required: +15 Vdc @ 25 mA
Possible Part Number and Supplier
Part Number: W50ATC
Supplier: Advanced Control Components
20 Meridian Road
Eatontown, NJ 07724
30 MHz to 2 GHz Low Noise Amplifier
Configuration
Two amplifiers are cascaded to obtain the necessary requirements. The specifications
listed below are the combined specifications of the cascaded amplifiers.
Critical Specifications
Frequency Range: 10 MHz to 2 GHz
Noise Figure: 1.8 dB
Gain: > 56 dB
1 dB Compression Minimum: +10 dBm
External Power Supply Required: +15 Vdc @ 300 mA
Possible Part Number and Supplier
Part Number: AFS3-00010300-18-5P-4 (2 required)
Supplier: MITEQ Inc.
100 Davids Drive
Hauppauge, NY 11788