beautypg.com

Burst modulation on/off ratio (esg-d only), Recommended equipment, Equipment setup – Atec Agilent-8563E User Manual

Page 82

background image

ESG Family Signal Generators

Performance Tests

17. Burst Modulation On/Off Ratio (ESG-D only)

Calibration Guide

4-33

17. Burst Modulation On/Off Ratio (ESG-D only)

This automated test verifies the burst modulation on/off ratio. The signal generator is
configured for an external burst input. The spectrum analyzer measures the RF output
power with and without the external burst applied. The power difference is the on/off ratio.

Recommended Equipment

• HP/Agilent 8563E Spectrum Analyzer

• HP/Agilent 8491A/B Option 010 Attenuator (10 dB)

• HP/Agilent 33120A Arbitrary Waveform Generator

Equipment Setup

Figure 4-30 Burst Modulation On/Off Ratio Test Setup