beautypg.com

Check for error messages, Frequency range and accuracy check – Atec Agilent-8563E User Manual

Page 5

background image

ESG Family Signal Generators

Operation Verification

Verification Procedures

Calibration Guide

2-3

2. Check for Error Messages

1. When the display is lit, check to see if the

ERR

annunciator is turned on.

2. If the

ERR

annunciator is turned on, review the error messages in the queue by pressing

Utility

>

Error Info

>

View Next Error Message

. The first error message in the queue will be

shown in the text area of the display. Refer to the error messages guide for information
about the error message.

If there is more than one error message (each message will be designated as 1 of n),
continue pressing the

View Next Error Message

softkey until you have seen all of the

messages.

3. When you have resolved all of the error messages, press

Clear Error Queue(s)

to delete the

messages. Cycle the power on the signal generator and then restart this procedure
(Check for Error Messages) at step 1.

3. Frequency Range and Accuracy Check

Frequency range is tested by determining the frequency accuracy relative to the timebase
at the frequency limits.

Connect the Test Equipment

Figure 2-1 Frequency Range and Accuracy Equipment Setup

Configure the Signal Generator

1. Preset the signal generator to normal (factory-defined) preset conditions.

2. Press

Mod On/Off

. The

MOD OFF

annunciator is displayed.

3. For frequencies < 10 MHz:

a. Press

Amplitude

. Enter 10 using the numeric keypad and press the

dBm

terminator

softkey.

b. Press

RF On/Off

. The

RF ON

annunciator is displayed.