Performance Tests

ESG Family Signal Generators

16. Pulse Modulation On/Off Ratio

4-32

Calibration Guide

16. Pulse Modulation On/Off Ratio

This automated test verifies the pulse modulation on/off ratio. The signal generator is
configured for an external pulse input. The spectrum analyzer measures the RF output
power with and without the external pulse applied. The power difference is the on/off ratio.

NOTE

This test does not test the high performance pulse circuitry used in
Option 1E6 instruments. See

“20. Pulse Rise/Fall Time (Option 1E6 only)”

on

page 4-36.

Recommended Equipment

• HP/Agilent 8563E Spectrum Analyzer

• HP/Agilent 8491A/B Option 010 Attenuator (10 dB)

• HP/Agilent 33120A Arbitrary Waveform Generator

Equipment Setup

Figure 4-29 Pulse Modulation On/Off Ratio Test Setup