Low Noise Floor Gain Adjust (ESG-AP and ESG-DP only)

5-32

Calibration Guide

Low Noise Floor Gain Adjust
(ESG-AP and ESG-DP only)

Description

This adjustment calibrates the gain adjust DAC on the output module for low noise floor
(LNF) mode. The instrument is set up with attenuation applied by way of the burst
modulator. This ensures that the RF chain is not in compression. The gain adjust DAC is
then set for a specific RF output power, as measured with a power meter.

Required Test Equipment

• HP/Agilent 438A Power Meter

• HP/Agilent 8482A Power Sensor

Procedure

Figure 5-28

LNF Gain Adjust Setup

1. Connect the equipment as shown above.

2. Preset all of the equipment.

3. Follow the instructions as they appear on the controller’s display.