Service Software

ESG Family Signal Generators

ESG Family Support Software Administration

3-8

Calibration Guide

1. Make sure that ESG_B is selected in the Select an Instrument Family list.

2. In the Select Model list, select the instrument model of the DUT to be adjusted or

tested.

3. In the Serial Number box, enter the complete serial number of the DUT.

4. In the Address box, enter the two-digit GPIB address of the DUT. (To display the

address on the signal generator, press

Utility

>

GPIB/RS-232

. The

GPIB Address

softkey

shows the current address.)

5. Select the OK button.

6. The Select Test Equipment and Tests window appears on the display. Close this

window.

Adding Test Equipment

Before any performance tests can be run, information regarding your specific test
equipment must be entered into the software’s database. This information includes the
serial numbers, calibration dates, GPIB address, and traceability data for each individual
component. Only equipment which has been entered into the system will be available for
the test equipment selection process during the test setup. To add test equipment, follow
the instructions below:

1. Refer to

Figure 3-2

. In the File drop-down menu, select Test Equipment.

Figure 3-2

Test Equipment in the File Drop-Down Menu

2. Refer to

Figure 3-3

. In the Test Equipment dialog box, select the Device Type for the

new equipment you are adding.