Adjustments

ESG Family Signal Generators

Modulation Source Relative Gain Compression Calibration

5-18

Calibration Guide

Modulation Source
Relative Gain Compression Calibration

Description

This test provides a scaling factor for all of the multiplexed FM modulation inputs. The
scaling factor is used by the signal generator’s firmware to scale the actual requested FM
deviation from the synthesizer board when the corresponding input is selected. Three
scaling factors (EXT1, EXT2, and INT1) are generated during this calibration. The
resulting values are used to calculate the calibration values that are then stored in the
signal generator’s firmware.

Required Test Equipment

• HP/Agilent 3458A Digital Multimeter

• HP/Agilent 53132A Option 050 Universal Counter

• HP/Agilent 33120A 15 MHz Function/Arbitrary Waveform Generator

Procedure

Figure 5-14

Modulation Source Relative Gain Compression Calibration Setup

1. Connect the equipment as shown above.

2. Preset all of the equipment.

3. Follow the instructions as they appear on the controller’s display.