ESG Family Signal Generators

Operation Verification

Verification Procedures

Calibration Guide

2-11

Table 2-2 Power Level Accuracy Setup 1 (Signal Generators without Option 1E6)

Frequency

Setting

Power Level

Setting

(dBm)

Lower Limit

(dBm)

Measured

Power

(dBm)

Upper Limit

(dBm)

277 kHz

+13

12.2

13.8

+7

6.5

7.5

0

0.5

0.5

5

5.5

4.5

15

15.5

14.5

2.516 MHz

+13

12.2

13.8

+7

6.5

7.5

0

0.5

0.5

5

5.5

4.5

15

15.5

14.5

270.1 MHz

+13

12.2

13.8

+7

6.5

7.5

0

0.5

0.5

5

5.5

4.5

15

−1

5.5

14.5

510.1 MHz

+13

12.2

13.8

+7

6.5

7.5

0

0.5

0.5

5

5.5

4.5

15

15.5

14.5

990.1 MHz

+13

12.2

13.8

+7

6.5

7.5

0

0.5

0.5

5

5.5

4.5

15

15.5

14.5

1350.1 MHz

+10

9.2

10.8

+7

6.5

7.5

0

0.5

0.5