Am accuracy check – Atec Agilent-8563E User Manual

Page 10

background image

Operation Verification

ESG Family Signal Generators

Verification Procedures

2-8

Calibration Guide

Configure the Measuring Receiver

1. Reset the measuring receiver.

2. Set the measuring receiver to FM mode.

3. Turn on Peak+ detector.

4. Turn on the 300 Hz high-pass filter.

5. Turn on the 3 kHz low-pass filter.

Measure the Deviations

1. Set the signal generator to the frequencies shown in

Table 2-5, “FM Accuracy,”

on

page 2-19.

2. Record the deviations measured and compare them to the limits shown in

Table 2-5

.

6. AM Accuracy Check

Connect the Test Equipment

Figure 2-5 AM Accuracy Equipment Setup

Configure the Signal Generator

1. Preset the signal generator to normal (factory-defined) preset conditions.

2. Press

AM

>

AM Off On

. The

AM

annunciator is displayed.

3. Press

AM Rate.

Enter 1 using the numeric keypad and press the

kHz

terminator softkey.

4. Press

Amplitude.

Enter 7 using the numeric keypad and press the

dBm

terminator

softkey.

5. Check that the

MOD ON

annunciator is displayed. If it is not, press

Mod On/Off

.

6. Press

RF On/Off

. The

RF ON

annunciator is displayed.