Operation Verification
ESG Family Signal Generators
Verification Procedures
2-8
Calibration Guide
Configure the Measuring Receiver
1. Reset the measuring receiver.
2. Set the measuring receiver to FM mode.
3. Turn on Peak+ detector.
4. Turn on the 300 Hz high-pass filter.
5. Turn on the 3 kHz low-pass filter.
Measure the Deviations
1. Set the signal generator to the frequencies shown in
2. Record the deviations measured and compare them to the limits shown in
6. AM Accuracy Check
Connect the Test Equipment
Figure 2-5 AM Accuracy Equipment Setup
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
AM
>
AM Off On
. The
AM
annunciator is displayed.
3. Press
AM Rate.
Enter 1 using the numeric keypad and press the
kHz
terminator softkey.
4. Press
Amplitude.
Enter 7 using the numeric keypad and press the
dBm
terminator
softkey.
5. Check that the
MOD ON
annunciator is displayed. If it is not, press
Mod On/Off
.
6. Press
RF On/Off
. The
RF ON
annunciator is displayed.